Initial Alignment Positions of AXAF High-Resolution Mirror Assembly (HRMA) during Calibration at the X-Ray Calibration Facility (XRCF).
Phase-IAP | Date, GMT | Pitch | Yaw | dX | |
Established | arcmin | arcmin | mm | Comments | |
1-IAP0 | 0.0 | 0.0 | 0.000 | Aligned X-ARM, HRMA fixed on blocks | |
1-IAP1 | 364 18:00 | 0.0 | 0.0 | 0.000 | Aligned X-ARM, original pre-vacuum IAP |
1-IAP2 | 001 05:00 | -1.1 | -1.3 | 0.000 | Initial failure IAP, calculated from encoder counts |
1-IAP3 | 001 12:15 | -1.5 | -0.5 | 0.000 | X-ray alignment test, change -0.4 pitch and +0.8 yaw wrt 1-IAP2 |
1-IAP4 | 002 18:22 | 0.0 | 0.0 | 0.000 | Aligned X-ARM |
1-IAP5 | 005 16:11 | 0.0 | -1.0 | 0.000 | X-ray quadrant-balance test, change -1.0 yaw wrt 1-IAP4 |
1-IAP6 | 026 08:08 | 0.0 | -1.0 | 0.000 | Same as 1-IAP5, but encoders zeroed with rack off |
1-IAP7 air | 027 20:00 | 0.0 | 0.0 | 0.000 | Aligned X-ARM air, GSE @ 70 F |
1-IAP7 | 028 15:01 | 0.0 | 0.0 | 0.000 | Aligned X-ARM, GSE @ 67 F |
1-IAP7a | 029 16:52 | 0.0 | 0.0 | 0.000 | Aligned X-ARM, GSE @ 55 F (Yaw 47 arcsec from 1-IAP7) |
1-IAP7b | 029 20:32 | 0.0 | 0.0 | 0.000 | Aligned X-ARM, GSE @ 54 F (Pitch 7 arcsec from 1-IAP7a) |
1-IAP8 | 036 18:00 | 0.0 | 0.0 | 0.000 | Aligned X-ARM, GSE @ 52.5 F (Encoders nonzero at home) |
2-IAP1 | 053 17:38 | 0.0 | 0.0 | 0.000 | Aligned X-ARM |
2-IAP2 | 058 23:16 | 0.0 | 0.0 | 0.203 | Aligned X-ARM, HRMA axial displacement when setting IAP |
2-IAP3 | 059 17:38 | 0.0 | 0.0 | 1.549 | Aligned X-ARM, HRMA axial displacement when setting IAP |
2-IAP4 | 064 20:38 | 0.0 | 0.0 | 0.076 | Aligned X-ARM, HRMA axial displacement when setting IAP |
2-IAP5 | 079 15:00 | 0.0 | 0.0 | 0.000 | Aligned X-ARM |
2-IAP6 | 091 05:15 | 0.0 | 0.0 | 0.000 | Aligned X-ARM, special IAP for ghost-image search |
Note a: | Column 1 gives Calibration phase hyphen Initial Alignment Position sequence number. | ||||
Note b: | Columns 3 and 4 give HRMA pitch and yaw relative to XRCF Alignment Reference Mirror (X-ARM). | ||||
Note c: | Column 5 gives axial error resulting from incorrect software algorithm for HRMA actuators. |